TitleProduct

Medical and Hospital Premature Incubator for New Born Baby

  • Price:

    Negotiable

  • minimum:

  • Total supply:

  • Delivery term:

    The date of payment from buyers deliver within days

  • seat:

    Jiangsu

  • Validity to:

    Long-term effective

  • Last update:

    2017-12-26 14:44

  • Browse the number:

    67

Send an inquiries

Company Profile

Zhangjiagang Yooda Medical Technology Co., Ltd.

By certification [File Integrity]

Contact:Mr. Carl(Mr.)  

Email:

Telephone:

Phone:

Area:Jiangsu

Address:Jiangsu

Website: http://yoodamed.sunla-tech.com/

Product Details


medical and hospital infant incubator for new born baby


Brief Introduction:
Two control modes: Air mode and baby mode controlled by micro-computer;
Set temperature, air temperature, baby temperature are displayed separately; Humidity concentration system
Inclination of bassinet is adjustable;
Failure alarm indication;
RS-232 connector;
Oxygen input interface;
Configured with fixed neonate bilirubin phototherapy equipment model: YDXHZ-90 with blue fluorescent lamp as the upper light source; And fixed neonate bilirubin phototherapy equipment model: YDXHZ-90L with LED blue light is optional.
Configured with fixed neonate bilirubin phototherapy equipment with SMD LED blue lights under the bassinet.
With DC motor
 
Standard Configuration:
Main body, mobile cabinet, temperature controller, baby compartment, I.V. Pole with shelf, Skin temperature sensor, mattress, fixed Neonate Bilirubin Phototherapy Equipment model: YDXHZ-90 with blue fluorescent lamp as the upper light source, fixed Neonate Bilirubin Phototherapy Equipment with SMD LED blue light under the bassinet.
 
Optional parts:
> 37 °C temperature override mode function, Vertical height adjustment stand(VHA stand), Weighing system, Neonate bilirubin phototherapy equipment model: YDXHZ-90L with LED blue light as upper light source and Disposable skin temperature sensor.
 
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz , 1000VA
Mode of control: Air mode and baby mode controlled by micro-computer
Control range of air temperature: 25°C -37°C (override mode: 37-39°C is optional)
Control range of baby temperature: 34-37°C (override mode: 37-38°C is optional)
Accuracy of skin temperature sensor: ±0.2°C
Variability of temperature≤0.5°C
Uniformity of temperature≤: 0.8°C (mattress is on the horizontal position)
Humidity display range: 0%RH 99%RH, Humidity control rang: 20% - 90%RH
Inclination of the bassinet: ±12°
Noise inside hood: 45dB(A) (Environment noise 35dB(A))
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment:
≥640pW/cm2(blue fluorescent lamp)
≥130011W//cm2(LED blue light)
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment under bassinet: • 800pW/cm2 (with SMD LED)
Built-in timer range: 0-9999 hrs 59min
Accuracy of weight display: ±1% (equipped weighing system)
 
Requirements of environment:
Environment temperature: 20 °C -30°C (Set temperature should be higher 3°C than the environment temperature)
Wind speed of environment: < 0.3m/s


Package
Package each unit is packed into two carton
Dimension main body 1085mm x 640mm x 910mm
cabinet 1125mm x 665mm x 725mm
Gross Weight(main body+cabinet) 78kg+68kg